finale bevestiging van ondersteuning word vereis. Sodra u bestelling geplaas is, wag asseblief vir finale bevestiging van die maatskappy. ( 97144476031)

Fluke 754 Documenting Process Calibrator-HART

0.00 AED
In stock
Product Details
UPC: 095969491754
Brand: Fluke
Voltage DC Range: 100.000 mV to 300.00 V
Voltage AC Range: 3.000 V (40 Hz to 500 Hz) / 0.001 V to 300.0 V (40 Hz to 500 Hz) / 0.1 V
Resistance Range: 10.000 Ω to 10.000 kΩ
Current DC Range: 30.000 mA to 110.00 mA
mA Source: 1 mA, 500 μA, 250 μA, 150 μA, 3 mA
Frequency: 1.00 to 110.00 Hz / 0.01 Hz to 11.01 to 50.00 kHz / 0.01 kHz
Loop Supply: Selectable, 26 V
Thermocouples: E, N, J, K, T, B, R, S, C, L, U, BP, XK
RTDs: Cu, Ni, Pt
Dimension: 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
Weight: 1.2 kg (2.7 lb)
Power: Internal Battery Pack Li-ion: 7.2V,4400mAh, 30 Wh
Operating temperature: -10°C to +50°C
Storage temperature: -20°C to +60°C
Dust/water resistance: Meets IP52, IEC 529

Fluke 754 Documenting Process Calibrator HART

The Fluke 754 Documenting Process Calibrator-HART is a cutting-edge instrument that offers a wide range of features, making it an essential tool for professionals in the process industry. This versatile calibrator is designed to calibrate and maintain HART and other instrumentation, providing users with the capability to measure, source, and simulate various process signals.

Key Features

The Fluke 754 Documenting Process Calibrator-HART is equipped with features that set it apart from other similar devices. It offers integrated HART communication, allowing for the calibration and configuration of HART instrumentation.
The calibrator can measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors and transmitters. It also sources/simulates volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters. The device's user-friendly interface and high-resolution display facilitate real-time data interpretation, while its robust data logging capabilities allow for detailed long-term analysis.

Applications

This versatile calibrator is well-suited for a diverse set of applications, including maintaining and calibrating process instrumentation. It can be effectively utilized in industries such as pharmaceuticals, energy, and petrochemicals to ensure the accurate and reliable operation of process equipment.
The Fluke 754 Documenting Process Calibrator-HART is designed to provide users with the data needed to calibrate and configure a wide range of instrumentation, ultimately leading to increased productivity and cost savings.
In conclusion, the Fluke 754 Documenting Process Calibrator-HART is a state-of-the-art device that significantly streamlines process calibration and maintenance. Its advanced features, wide-ranging applications, and availability make it an indispensable tool for professionals in the process industry.

Comparing the Fluke 754

With Other Calibration Tools

When compared to other calibration tools, the Fluke 754 distinguishes itself through its advanced features, rugged design, and integrated HART communication, offering superior value and functionality.

Cost-Effectiveness and ROI

Despite its premium features, the Fluke 754 proves to be a cost-effective solution in the long run, owing to its durability, efficiency, and the time saved in calibration processes.

Fluke 754 Process Calibrator Specifications:

Measurement Accuracy
Voltage DC Range/ Resolution 1 Year 2 Years
100.000 mV 0.02% + 0.005 mV 0.03% + 0.005 mV
3.00000 V 0.02% + 0.00005 V 0.03% + 0.00005 V
30.0000 V 0.02% + 0.0005 V 0.03% + 0.0005 V
300.00 V 0.05% + 0.05 V 0.07% + 0.05 V
Voltage AC 3.000 V (40 Hz to 500 Hz) / 0.001 V 0.5% + 0.002 V 1.0% + 0.004 V
30.00 V (40 Hz to 500 Hz) / 0.01 V 0.5% + 0.02 V 1.0% + 0.04 V
300.0 V (40 Hz to 500 Hz) / 0.1 V 0.5% + 0.2 V 1.0% + 0.2 V
Current DC 30.000 mA 0.01% + 5 uA 0.015% + 7 uA
110.00 mA 0.01% + 20 uA 0.015% + 30 uA
Resistance 10.000 Ω 0.05% + 50 mΩ 0.07% + 70 mΩ
100.00 Ω 0.05% + 50 mΩ 0.07% + 70 mΩ
1.0000 kΩ 0.05% + 500 mΩ 0.07% + 0.5 Ω
10.000 kΩ 0.1% + 10 Ω 0.15% + 15 Ω
Frequency 1.00 to 110.00 Hz / 0.01 Hz 0.05 Hz
110.1 to 1100.0 Hz / 0.1 Hz 0.5 Hz
1.101 to 11.000 kHz / 0.001 kHz 0.005 kHz
11.01 to 50.00 kHz / 0.01 kHz 0.05 kHz
Source Accuracy
1 Year 2 Years
Voltage DC 100.000 mV 0.01% + 0.005 mV 0.015% + 0.005 mV
1.00000 V 0.01% + 0.00005 V 0.015% + 0.0005 V
15.0000 V 0.01% + 0.0005 V 0.015% + 0.0005 V
Current DC 22.000 mA (source) 0.01% + 0.003 mA 0.02% + 0.003 mA
Current sink (simulate) 0.02% + 0.007 mA 0.04% + 0.007 mA
Resistance 10.000 Ω 0.01% + 10 mΩ 0.015% + 15 mΩ
100.00 Ω 0.01% + 20 mΩ 0.015% + 30 mΩ
1.0000 kΩ 0.02% + 0.2 Ω 0.03% + 0.3 Ω
10.000 kΩ 0.02% + 3 Ω 0.03% + 5 Ω
Frequency 0.1 to 10.99 Hz 0.01 Hz
0.01 to 10.99 Hz 0.01 Hz
11.00 to 109.99 Hz 0.1 Hz
110.0 to 1099.9 Hz 0.1 Hz
1.100 to 21.999 kHz 0.002 kHz
22.000 to 50.000 kHz 0.005 kHz
Technical Data
Data log functions Measure functions Voltage, current, resistance, frequency, temperature, pressure
Reading rate 1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length 8000 readings (7980 for 30 or 60 readings/minute)
Ramp functions Source functions Voltage, current, resistance, frequency, temperature
Rate 4 steps/second
Trip detect Continuity or voltage (continuity detection not available when sourcing current)
Loop power function Voltage Selectable, 26 V
Accuracy 10%, 18 V minimum at 22 mA
Maximum current 25 mA, short circuit protected
Maximum input voltage 50 V DC
Step functions Source functions Voltage, current, resistance, frequency, temperature
Manual step Selectable step, change with arrow buttons
Autostep Fully programmable for function, start delay, stepvalue, time per step, repeat
Environmental Specifications
Operating temperature -10°C to +50°C
Storage temperature -20°C to +60°C
Dust/water resistance Meets IP52, IEC 529
Operating altitude 3000 m above mean sea level (9842 ft)
Safety Specifications
Agency approvals CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
Mechanical and General Specifications
Size 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
Weight 1.2 kg (2.7 lb)
Batteries Internal Battery Pack Li-ion: 7.2V,4400mAh, 30 Wh
Battery life >8 hours typical
Battery replacement Replace without opening calibrator; no tools required
Side port connections Pressure module connector
USB Connector to interface to your PC
Digital instrument (HART) connector
Connection for optional battery charger/eliminator
Data storage capacity 1 week of calibration procedures results
90 day specifications The standard specification interval for the 750 Seriesare 1 and 2 years.
Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2.
Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant.
Save this product for later
Share this product with your friends
Fluke 754 Documenting Process Calibrator-HART