مطلوب تأكيد نهائي من الدعم. بمجرد تقديم طلبك، يرجى انتظار التأكيد النهائي من الشركة. ( 97144476031)
Fluke 719 100G Pressure Calibrator
0.00 AED
متوفر في المتجر
تفاصيل المنتج
Brand:
Fluke
Range:
-12 to 120 psi, -850 mbar to 8 bar
Resolution:
0.01 psi, 1 mbar
Accuracy:
0.025% range for 6 months, 0.035% of range for one (1) year
Operating Temp.:
-10°C to 55°C
Storage Temp.:
-30°C to 60°C
Size:
60 x 87 x 210 mm (2.19 x 3.41 x 8.28 in)
Weight:
912 g (2 lb)
Loop Power Range:
24 V DC
Shock & Vibration:
1 meter drop test, Random, 2 G, 5-500 Hz
Fluke 719 100G Pressure Calibrator
The Fluke 719 100G Pressure Calibrator is a versatile instrument designed for precision pressure calibration in industrial settings. Let’s delve into its features, applications, and specifications:
Key Features
- Built-in Electric Pump: The Fluke 719 boasts a convenient built-in electric pump, allowing one-handed pressure calibration. Say goodbye to manual hand pumping!
- Pressure Measurement Uncertainty: With an impressive uncertainty of 0.025%, it ensures accurate pressure readings.
- mA Measurement and Sourcing: The calibrator can measure and source mA signals with an exceptional accuracy of 0.015%, making it ideal for testing valves and I/P (current-to-pressure) devices.
- Pressure Ranges: It calibrates both 30 psi and 100 psi pressure ranges.
- Precision Adjustment Vernier: The vernier facilitates easy and precise pressure calibration adjustments.
- Transmitter Power: The Fluke 719 can power transmitters during testing using a 24 V loop supply.
- Three-Year Warranty: Rest assured with the included warranty.
Applications and Utilizations
- Calibrating P/I Transmitters: The Fluke 719 ensures accurate calibration of pressure-to-current (P/I) transmitters.
- I/P Devices Calibration: It’s equally
Fluke 719 100G Specifications
General Specifications
- Storage temperature: -30°C to 60°C
- Operating temperature: -10°C to 55°C
- Relative humidity:
- 95% (-10°C to 30°C)
- 75% (30°C to 40°C):
- 45% (40°C to 50°C):
- 35% (50°C to 55°C):
- Shock: 1 meter drop test
- Safety: CSA C22.2 No. 1010.1:1992
- EMC: EN50082-1:1992 and EN55022:1994 Class B
- Size (H x W x L): 60 x 87 x 210 mm (2.19 x 3.41 x 8.28 in)
- Weight: 912 g (2 lb)
- Battery: Two 9 V alkaline batteries
- Battery life: 12 hours with 12 mA into 500 Ω, 200 pump cycles to 100 psi, 500 pump cycles to 30 psi
- Warranty: Three-years, one year for pump
Specifications
- Fluke-719 100G:
- Range: -12 to 120 psi, -850 mbar to 8 bar
- Resolution: 0.01 psi, 1 mbar
- Accuracy:
- 0.025% range for 6 months
- 0.035% of range for one (1) year
- mA:
- Range: 0 to 24 mA
- Resolution: 0.001 mA
- Accuracy: .015% Rdg + 2 counts
- Max load, 1000 Ω, 750 Ω ohm in HART mode:
- Loop supply:
- Range: 24 V DC
- Resolution: N/A
- Accuracy: ±10%
Pump Specifications
- Maximum pressure generation: 120 psi (8 BAR) (100 psi model)
- Pump to:
- -30 psi in less than 15 seconds1
- -100 psi in less than 45 seconds1
- 1. With 1 meter of 1/8 inch test hose
- Pump cycles between rebuilds: 5000
- Total pump life: 20,000 calibration cycle
حفظ هذا المنتج في وقت لاحق
Fluke 719 100G Pressure Calibrator
عرض الأسعار في:
AED