final confirmation from support is required. Once your order is placed, please await final confirmation from the company. (+97144476031)
Eurotron PLB 250 Portable Liquid Bath Calibrator
0.00 AED
In stock
Product Details
Eurotron PLB 250 Portable Liquid Bath Calibrator
The Eurotron PLB 250 Portable Liquid Bath Calibrator is an exceptional tool for professionals who require precise and reliable temperature calibration.
It features a robust and compact design, making it highly portable and easy to transport between job sites. Its sturdy construction ensures durability, capable of withstanding the rigors of fieldwork. The user-friendly interface and clear digital display simplify the calibration process, catering to both novice and experienced users.
Eurotron PLB 250 Key Features
- Temperature range:RT to 250 ºC
- Large calibration volume: Ø 60 x 170 mm
- Accuracy better than ±0.08 C using external reference sensor
- High stability: ± 0.02 ºC
- Best in class radial uniformity: ± 0.02 ºC
- Inputs for reference sensor and sensor under test
- Fully programmable ramp
- Automatic switch test function
- Use as stirred liquid bath, ice point, dry block, black body target and surface sensors temperature calibrator
- RS232 and calibration software as standard
Eurotron PLB 250 Specifications
Temperature Range | |
@ 20ºC | RT to 250 °C |
Stability | |
As Liquid Bath | ± 0.01 °C |
As Dry block | ± 0.01 °C |
Uniformity as Liquid Bath | |
Axial | ± 0.02 °C @ -10 °C ± 0.02 °C@ 0°C ± 0.01 °C @ 120 °C |
Radial | ± 0.01 °C @ -10 °C ± 0.01 °C@ 0°C ± 0.02 °C @ 120 °C |
Accuracy | |
with Internal Sensor | ± 0.15 °C |
with External Sensor SRS (*) | ± 0.08 °C ( TC version only) |
Display | |
Resolution | user-selectable 0.1 , 0.01 C |
Units | user-selectable C, F,K |
Calibration Volume | |
As Liquid Bath | 54 x 150 mm |
As dry block | 60 x 170 mm |
Heating time | 20 to 250°C: 30 min ( include time to stability) |
Cooling time | 250 to 60°C: 60 min (include time to stability) |
Thermocouple Input (TC version) | |
Type | K,N,R,S,J,E |
Range | 0 to 1300 °C |
Accuracy | ± 1°C Include CJC error |
Resolution | user-selectable 0.1 , 0.01 °C |
RTD’S input ( TC version only) | |
Type | Pt 100 (3/4 wires) |
Range | -100 C to 700 °C |
Accuracy | ± 0.3 °C |
Resolution | user-selectable 0.1 , 0.01 °C |
General | |
Computer interface | RS 232 as standard ( UBS adapter supplied as option) and EIUK software included |
Switch Test | Yes |
Ramp | Yes fully programmable |
Weight | 8 Kg |
Dimensions | 160x340x330 mm |
Power Requirement | 115 o 230 V – 50/60 Hz |
Power | 500 VA |
Notes | * Require SRS sensor and special UKAS matching calibration. Long-term drift not included |