final confirmation from support is required. Once your order is placed, please await final confirmation from the company. (+97144476031)

Fluke 753 Documenting Process Calibrator

0.00 AED
In stock
Product Details
Brand: Fluke
Voltage DC Range: 100.000 mV to 300.00 V
Voltage AC Range: 3.000 V (40 Hz to 500 Hz) / 0.001 V to 300.0 V (40 Hz to 500 Hz) / 0.1 V
Resistance Range: 10.000 Ω to 10.000 kΩ
Current DC Range: 30.000 mA to 110.00 mA
mA Source: 22.000 mA (source)
mV: 100.000 mV
Loop Supply: Selectable, 26 V
Thermocouples: E, N, J, K, T, B, R, S, C, L, U, BP, XK
RTDs: Cu, Ni, Pt
Dimension: 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
Weight: 1.2 kg (2.7 lb)
Power: Internal Battery Pack Li-ion: 7.2V,4400mAh, 30 Wh
Operating temperature: -10°C to +50°C
Storage temperature: -20°C to +60°C
Dust/water resistance: Meets IP52, IEC 529

Fluke 753 Documenting Process Calibrator

The Fluke 753 Documenting Process Calibrator is a powerful and multifunctional tool used for calibration and measurement across various industries. This calibrator allows users to download procedures, lists, and instructions created with software, and upload data for printing, archiving, and analysis. It is suitable for measuring and sourcing a wide range of parameters, including voltage, current, temperature, and pressure.

Key Features

  • The Fluke 753 Documenting Process Calibrator offers a range of features that make it a valuable instrument for professionals in the field. Some of its key features include:
  • Multifunctionality: The calibrator is capable of measuring and sourcing various parameters, making it a versatile tool for different calibration needs.
  • Documentation: Users can download and upload procedures, lists, and instructions, allowing for efficient and organized calibration processes.
  • Software Compatibility: The calibrator is compatible with software for creating and managing calibration procedures, adding to its flexibility and usability.
Applications
The Fluke 753 Documenting Process Calibrator finds application in diverse industries, including manufacturing, petrochemical, pharmaceutical, and power generation. Its ability to calibrate and measure different parameters makes it an essential tool for ensuring the accuracy and reliability of process instruments and control systems.
Benefits
The use of the Fluke 753 Documenting Process Calibrator offers several benefits, including:
  • Efficiency: The ability to download and upload procedures and data streamlines the calibration process, saving time and effort.
  • Accuracy: The calibrator's multifunctionality and precision contribute to accurate measurement and calibration results.
  • Compliance: By using a documented calibration process, organizations can ensure compliance with quality and regulatory standards.
In conclusion, the Fluke 753 Documenting Process Calibrator is a valuable instrument for professionals involved in calibration and measurement activities. Its multifunctionality, documentation capabilities, and compatibility with software make it a versatile and efficient tool for a wide range of applications and industries.

Fluke 753 price

The Fluke 753 allows users to automate calibration procedures and document the results efficiently, ensuring compliance with standards such as ISO 9000, FDA, EPA, and OSHA regulations. It features a user-friendly interface with a graphical display, a USB port for easy data transfer, and a long-lasting Li-ion battery.

Fluke 753 Specifications

Measurement Accuracy
Voltage DC Range /Resolution 1 Year 2 Years
100.000 mV 0.02% + 0.005 mV 0.03% + 0.005 mV
3.00000 V 0.02% + 0.00005 V 0.03% + 0.00005 V
30.0000 V 0.02% + 0.0005 V 0.03% + 0.0005 V
300.00 V 0.05% + 0.05 V 0.07% + 0.05 V
Voltage AC 3.000 V (40 Hz to 500 Hz) / 0.001 V 0.5% + 0.002 V 1.0% + 0.004 V
30.00 V (40 Hz to 500 Hz) / 0.01 V 0.5% + 0.02 V 1.0% + 0.04 V
300.0 V (40 Hz to 500 Hz) / 0.1 V 0.5% + 0.2 V 1.0% + 0.2 V
Current DC 30.000 mA 0.01% + 5 uA 0.015% + 7 uA
110.00 mA 0.01% + 20 uA 0.015% + 30 uA
Resistance 10.000 Ω 0.05% + 50 mΩ 0.07% + 70 mΩ
100.00 Ω 0.05% + 50 mΩ 0.07% + 70 mΩ
1.0000 kΩ 0.05% + 500 mΩ 0.07% + 0.5 Ω
10.000 kΩ 0.1% + 10 Ω 0.15% + 15 Ω
Frequency 1.00 to 110.00 Hz / 0.01 Hz 0.05 Hz
110.1 to 1100.0 Hz / 0.1 Hz 0.5 Hz
1.101 to 11.000 kHz / 0.001 kHz 0.005 kHz
11.01 to 50.00 kHz / 0.01 kHz 0.05 kHz
Source Accuracy
1 Year 2 Years
Voltage DC 100.000 mV 0.01% + 0.005 mV 0.015% + 0.005 mV
1.00000 V 0.01% + 0.00005 V 0.015% + 0.0005 V
15.0000 V 0.01% + 0.0005 V 0.015% + 0.0005 V
Current DC 22.000 mA (source) 0.01% + 0.003 mA 0.02% + 0.003 mA
Current sink (simulate) 0.02% + 0.007 mA 0.04% + 0.007 mA
Resistance 10.000 Ω 0.01% + 10 mΩ 0.015% + 15 mΩ
100.00 Ω 0.01% + 20 mΩ 0.015% + 30 mΩ
1.0000 kΩ 0.02% + 0.2 Ω 0.03% + 0.3 Ω
10.000 kΩ 0.02% + 3 Ω 0.03% + 5 Ω
Frequency 0.1 to 10.99 Hz 0.01 Hz
0.01 to 10.99 Hz 0.01 Hz
11.00 to 109.99 Hz 0.1 Hz
110.0 to 1099.9 Hz 0.1 Hz
1.100 to 21.999 kHz 0.002 kHz
22.000 to 50.000 kHz 0.005 kHz
Technical Data
Data log functions Measure functions Voltage, current, resistance, frequency, temperature, pressure
Reading rate 1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length 8000 readings (7980 for 30 or 60 readings/minute)
Ramp functions Source functions Voltage, current, resistance, frequency, temperature
Rate 4 steps/second
Trip detect Continuity or voltage (continuity detection not available when sourcing current)
Loop power function Voltage Selectable, 26 V
Accuracy 10%, 18 V minimum at 22 mA
Maximum current 25 mA, short circuit protected
Maximum input voltage 50 V DC
Step functions Source functions Voltage, current, resistance, frequency, temperature
Manual step Selectable step, change with arrow buttons
Autostep Fully programmable for function, start delay, stepvalue, time per step, repeat
Environmental Specifications
Operating temperature -10°C to +50°C
Storage temperature -20°C to +60°C
Dust/water resistance Meets IP52, IEC 529
Operating altitude 3000 m above mean sea level (9842 ft)
Safety Specifications
Agency approvals CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
Mechanical and General Specifications
Size 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
Weight 1.2 kg (2.7 lb)
Batteries Internal Battery Pack Li-ion: 7.2 V, 4400 mAh, 30 Wh
Battery life > 8 hours typical
Battery replacement Via snap-shut door without opening calibrator; no tools required
Side port connections Pressure module connector
USB Connector to interface to your PC
Connection for optional battery charger/eliminator
Data storage capacity 1 week of calibration procedures results
90 day specifications The standard specification interval for the 750 Series are 1 and 2 years.
Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2.
Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant.
Save this product for later
Share this product with your friends
Fluke 753 Documenting Process Calibrator