final confirmation from support is required. Once your order is placed, please await final confirmation from the company. (+97144476031)
Tektronix AFG31151 Arbitrary Function Generator
0.00 AED
In stock
Product Details
Brand:
Tektronix
Channels :
1
Sine Frequency Range :
1 µHz to 150 MHz
Sample Rate :
2 GSa/s (16,384 pts)
Vertical Resolution :
14 bits
Amplitude Range (50 Ω) :
≤ 60 MHz: 1 mVP-P to 10 VP-P
Max Frequency Output :
150 MHz
Waveform Memory Size :
16 Mpts (128 Mpts optional)
Rise/Fall Time (Typical) :
≤ 3.0 ns (Amplitude ≤ 5Vpp)
Modulation:
AM, FM, PM
Output impedance:
50 Ω
Load impedance setting:
Selectable: 50 Ω, 1 Ω to 10.0 kΩ, High Z (Adjusts displayed amplitude according to selected load impedance)
Isolation:
42 Vpk maximum to earth ground
Short-circuit protection:
Signal outputs are robust against permanent shorts against floating ground
Run modes:
Continuous, Modulation, Sweep and Burst
Standard waveforms:
Sine, Square, Pulse, Ramp, More (Noise, DC,Sin(x)/x, Gaussian, Lorentz, Exponential Rise, Exponential Decay, Haversine )
Arbitrary waveforms:
Sampling clock: 150 MSa/s, 2 GSa/s (model and waveform length apply), Vertical resolution: 14 bits,
Frequency range:
Continuous mode: 1 μHz to 150 MHz and Burst mode: 1 μHz to 75 MHz
Dimensions:
191.8 mm x 412.8 mm x 143.3 mm
Weight:
4.7 kg (10.4 lb.)
Operating temperature:
0 °C to +50 °C (32 °F to 122 °F)
Storage temperature:
-30 °C to +70 °C (-22 °F to 158 °F)
Altitude:
Up to 3,000 m (9,842 ft.)
Safety:
UL 61010-1:2004, CAN/CSA C22.2 No. 61010-1:2004, IEC 61010-1:2001
Display:
9-inch capacitive touch screen
Tektronix AFG31151
The Tektronix AFG31151 Arbitrary Function Generator is a versatile and high-performance tool designed to meet the needs of engineers and researchers across various fields. This device is part of the AFG31000 series, which is known for its advanced capabilities and user-friendly interface.
One of the standout features of the AFG31151 is its ability to generate both traditional and arbitrary waveforms. This flexibility allows users to replicate real-world signals and perform a wide range of tests and measurements. The built-in arbitrary waveform generation capability means that users can create complex waveforms directly on the device, without needing additional software or hardware.
The AFG31151 also includes real-time waveform monitoring, which is facilitated by its large touchscreen display. This feature allows users to view and verify the actual waveform at the device under test (DUT) in real-time, eliminating the need for an oscilloscope and reducing the risk of mismatched impedance affecting test results. This real-time monitoring is made possible by Tektronix’s patented InstaView™ technology.
Another notable aspect of the AFG31151 is its intuitive user interface. The large, capacitive touchscreen operates much like a modern smartphone, making it easy to navigate through settings and options. This ease of use is particularly beneficial in complex testing environments where quick adjustments and real-time feedback are crucial.
The AFG31151 is also equipped with advanced modes for more sophisticated testing needs. For instance, the double pulse test feature allows users to generate two waveforms with varying pulse widths directly on the touchscreen, which is particularly useful for evaluating the dynamic behaviors of power devices like MOSFETs and IGBTs.
Overall, the Tektronix AFG31151 Arbitrary Function Generator is a powerful and flexible tool that supports a wide range of applications, from basic signal generation to advanced research and development. Its combination of advanced features, user-friendly interface, and real-time monitoring capabilities make it a valuable asset for any laboratory or testing environment.
Tektronix AFG31151 Specifications
Sine frequency range | 150 MHz |
Number of channels | 1 |
Range (into 50 Ω) | ≤ 200 MHz: 1 mVP-P to 5`VP-P |
> 200 MHz to ≤ 250 MHz: 1 mVP-P to 4`VP-P | |
Range (into open circuit or High-Z) | ≤ 200 MHz: 2 mVP-P to 10`VP-P |
> 200 MHz to ≤ 250 MHz: 2 mVP-P to 8`VP-P | |
Accuracy | ± (1% of setting +1 mVP-P) |
(1 kHz sine, 0 V offset, amplitude > 1 mVP-P) | |
Resolution | 0.1 mVP-P, 0.1`mVRMS, 1`mV, 0.1`dBm or 4`digits |
Units | Vpp, Vrms (excluding Arb and Noise), dBm (sine wave only), Volt (High Level and Low Level) |
Offset | |
Range (into 50 Ω) | ±(2.5 VPK - AmplitudeP-P ÷ 2) |
Range (into open circuit or High-Z) | ±(5 VPK - AmplitudeP-P ÷ 2) |
Accuracy | ± (1% of |setting| +1 mV + 0.5% of Amplitude (VP-P)) |
Resolution | 1 mV or 4 digits |
Output impedance
|
50 Ω |
Load impedance setting
|
Selectable: 50 Ω, 1 Ω to 10.0 kΩ, High Z (Adjusts displayed amplitude according to selected load impedance) |
Isolation
|
42 Vpk maximum to earth ground |
Short-circuit protection
|
Signal outputs are robust against permanent shorts against floating ground |
Overcurrent protection
|
When incoming current is greater than 250 mA, the output channels are protected with relays that disconnect the AFG from the device under test. Connection can be resumed by user after removing the incoming current |
Run modes | Continuous, Modulation, Sweep and Burst |
Standard waveforms | Sine, Square, Pulse, Ramp, More (Noise, DC,Sin(x)/x, Gaussian, Lorentz, Exponential Rise, Exponential Decay, Haversine ) |
Arbitrary waveforms | Sampling clock: 250 MSa/s, 1 GSa/s or 2 GSa/s (model and waveform length apply) |
Vertical resolution: 14 bits | |
Waveform length: | 2 to 131,072 points |
Frequency range | |
Continuous mode | 1 μHz to 150 MHz |
Burst mode | 1 μHz to 75 MHz |
Physical characteristics | |
Dimensions
|
|
Height | 191.8 mm (7.55 in.) |
Width | 412.8 mm (16.25 in.) |
Depth | 143.3 mm (5.64 in.) |
Weight
|
4.7 kg (10.4 lb.) |
EMC, environment, and safety | |
Operating Temperature
|
0 °C to +50 °C (32 °F to 122 °F) |
Nonoperating Temperature | -30 °C to +70 °C (-22 °F to 158 °F) |
Humidity
|
|
Operating | ≤ 80%, 0 °C to 40 °C (32 °F to104 °F) |
≤ 60%, > 40°C to 50°C (104 °F to 122 °F), noncondensing | |
Nonoperating | 5% to 90%, < 40 °C (< 104 °F), noncondensing |
5% to 80%, ≥ 40 °C to 60 °C (≥ 104 °F to 140 °F), noncondensing | |
5% to 40%, > 60 °C to 70 °C (> 140 °F to 158 °F), noncondensing | |
Altitude
|
|
Operating | Up to 3,000 m (9,842 ft.) |
Nonoperating | Up to 12,000 m (39,370 ft.) |
EMC compliance
|
EN61326-1:2013, EN 61326-2-1:2013 |
European Union | EU Council Directive 2004/108/EC |
Safety
|
UL 61010-1:2004 |
CAN/CSA C22.2 No. 61010-1:2004 | IEC 61010-1:2001 |
Over-temperature protection
|
Instrument is protected from over-temperature by turning off outputs |
Save this product for later
Tektronix AFG31151 Arbitrary Function Generator
Display prices in:
AED